Rob Sluys wrote on Tuesday, April 22, 1997 2:53 PM >> I see that you're supposed to have one of these flow meter things... = I'm wondering if they actually serve a purpose (like to not overpower the sensor) or are they just to connect = the inflator hose with the tubing << As you know, the oxygen sensor reacts to partial pressure, not fraction, = of oxygen. The pressure measured at the sensor can only be converted to = an equivalent fraction, if the flow rate is within recommended limits. = Too high of a flow can fool the sensor to read a higher partial pressure = and in return display a wrong fraction. Andy * Direct any flames and personal attacks to: AndyInsult@ao*.co* * -- Send mail for the `techdiver' mailing list to `techdiver@aquanaut.com'. Send list subscription requests to `techdiver-request@aquanaut.com'.